PLAT910 Type_3 Test for number of missingreflections below Theta(Min)
Possible causes:Experimental beamstop Theta(min) limit set too high; A large
unit-cell causingreflections effected by the beamstop. A possible technical
solution on CCD based equipmentinvolves the collection of additional images
with the detector at alarger distance from the crystal with the beamstop
setting changedaccordingly. Alternatively, a large number of low-order
'outlier' reflections mighthave been 'Omitted' deliberately from the (final)
least-squares refinement.The test is against the actual Theta(Min) of the
reflections in the *.fcffile. See the *.ckf file for details of the missing
reflections. Good qualitylow order reflections might be relevant for SQUEEZE
and similar solventmodelling refinement techniques.
Example reply:
Author response:
Those reflections are effected by the beamstop and the detector gap.