PLAT029_ALERT_3_A_diffrn_measured_fraction_theta_full value Low . 0.896 Why?
PLAT029 Type_3 Check the reported_diffrn_measured_fraction_theta_full
Ideally (and a requirementfor publication in Acta Crystallographica),
this fraction should beclose to 1.0 for theta-full greater or equal
to sin(theta/lambda) = 0.6(i.e. 25.24 degrees for MoKa and 67.7 degrees
for CuKa radiation).
The three major causes ofincomplete data sets are:
1 - A missing cusp of datadue to data collection by rotation around
the spindle axis only(standard on some image-plate systems).
Cure: collect anadditional data set after remounting the crystal.
2 - The DENZOimageprocessing package has problems with certain strong
reflections. They areoften excluded from the data set.
Cure: Add an additionalscan at lower power setting in order to
include strong loworder reflections.
3 - Incomplete scans,possibly based on erroneously assumed higher than
actual symmetry.
Note: The default value of_diffrn_measured_fraction_theta_full that
is automatically calculatedand inserted in the CIF by SHELXL-97 might
generate A-level ALERTSwhen significant numbers of reflections are
missing at higher thetavalues. In order to avoid such an ALERT,
substitute the valuescalculated with the SHELXL instruction 'ACTA 50'
for _diffrn_reflns_theta_fulland _diffrn_measured_fraction_theta_full
respectively. ForMo-radiation, corresponding values of 25 degees (or
higher) and 0.99 (orhigher) are expected. (See SHELXL manual).
PLATON may be used toanalyse the case at hand (by invoking either the
'FCF-VALIDATION' mode orthe 'ASYM-VIEW' mode).